RFQ/OBP400/2025/26

Annual Calibration for Orbital platform shaker Model 262
RFQ/OBP400/2025/26
Calibration for Orbishaker to ensure accurate performance of the shaker rotation speed RPM timer and load capacity for consistent operation Calibration Parameters 1 Rotational speed RPM Range Verify accuracy across the full range e.g. 10-250 RPM using a calibrated tachometer or strobe light Tolerance 5% of set value Industry standard for orbital shakers 2 Timer Functions Accuracy Validate countdown timer up to 120 Minutes against a calibrated stopwatch ensuring automatic halt at Zero 3 Load Capacity Test Confirm stability and performance under maximum load e.g 30KG to ensure no deviation in RPM or platform balance 4 Compliance and Documentation Certification Provide a calibration certification traceable to national standards eg SANAS
Request for Quote
2026/02/26 16:00:00
No