RFQ/OBP375/2025/26
Purchasing of CO2 incubator for tissue culture
RFQ/OBP375/2025/26
Category Specification Requirement Capacity Minimum 200 L internal chamber volume (acceptable range 200–300 L). Temperature Control Temperature range: 5°C above ambient to at least 60°C; Uniformity ±0.2°C; Stability ±0.1°C. Heating System Direct heat or air-jacketed system ensuring fast recovery after door opening. CO₂ Control Adjustable CO₂ concentration 0–20%; Accuracy ±0.1%; Automatic regulation via microprocessor/PID system. CO₂ Sensor Technology Infrared (IR) CO₂ sensor (heat-resistant type preferred). O₂ Control (Optional) O₂ suppression/control (1–19% range) where applicable. Humidity System Relative humidity ≥ 95% with integrated water reservoir or humidity pan. Sterilisation / Decontamination Built-in 90°C moist heat sterilisation cycle or 180°C dry-heat high-temperature sterilisation capability. Filtration HEPA or ULPA filtration system for ISO Class 5 internal air quality. Inner Chamber Material Stainless steel (304/316) or solid copper interior (as applicable). Exterior Finish Powder-coated steel, corrosion-resistant, antimicrobial coating preferred Doors Outer solid door + inner glass door or multiple inner doors for parameter stability. Shelving Minimum 3 adjustable stainless steel shelves, expandable. Recovery Time Rapid temperature, CO₂, and humidity recovery after door openings (≤5 minutes typical). Decontamination Safety Automatic lockout during sterilisation cycle; temperature and sensor protection. Safety Features Over-temperature protection, CO₂ over-injection alarms, sensor fault alarms. Controls & Display Digital controller with touch panel; real-time display of CO₂, temperature, humidity; alarm logs. Data Logging USB/RS-232/Ethernet port or internal memory for data recording. Gas Inlet Requirements CO₂ inlet with 0.3–0.7 MPa rating; in-line CO₂ filter included. Power Requirements Standard: 220–240 V, 50 Hz; Max power consumption ≤1.2 kW. Optional Accessories CO₂ gas regulator, N₂ backup, extra shelves, UV lamp, cooling system, stand or stacking frame. Cooling System (Optional) Ability to operate below ambient temperature (e.g., 8°C below ambient) if cooling model is required. Include IQ/OQ
Request for Quote
2026/02/17 15:00:00
No